Abstract
Purpose
Ultrasound (US)-guided percutaneous puncture technology can realize real-time, minimally invasive interventional therapy without radiation. The location accuracy of the puncture needle directly determines the precision and safety of the operation. It is a challenge for novices and young surgeons to perform a free-hand puncture guided by the ultrasound images to achieve the desired accuracy. This work aims to develop a robotic system to assist surgeons to perform percutaneous punctures with high precision.
Methods
An US-guided puncture robot was designed to allow the mounting and control of the needle to achieve localization and insertion. The US probe fitted within the puncture robot was held by a passive arm. Moreover, the puncture robot was calibrated with a novel calibration method to achieve coordinate transformation between the robot and the US image. The system allowed the operators to plan the puncture target and puncture path on US images, and the robot performed needle insertion automatically. Five groups of puncture experiments were performed to verify the validity and accuracy of the proposed robotic system.
Results
Assisted by the robotic system, the positioning and orientation accuracies of the needle insertion were 0.9 ± 0.29 mm and 0.76 ± 0.34°, respectively. These are improved compared with the results obtained with the free-hand puncture (1.82 ± 0.51 mm and 2.79 ± 1.32°, respectively). Moreover, the proposed robotic system can reduce the operation time and number of needle insertions (14.28 ± 3.21 s and one needle insertion, respectively), compared with the free-hand puncture (25.14 ± 6.09 s and 1.96 ± 0.68 needle insertions, respectively).
Conclusion
A robotic system for percutaneous puncture guided by US images was developed and demonstrated. The experimental results indicate that the proposed system is accurate and feasible. It can assist novices and young surgeons to perform the puncture operation with increased accuracy.










Similar content being viewed by others
References
Soni NJ, Franco-Sadud R, Schnobrich D, Dancel R, Tierney DM, Salame G, Restrepo MI, McHardy P (2016) Ultrasound guidance for lumbar puncture. Neurol Clin Pract 6(4):358–368
Mahmoud MZ, Aslam M, Alsaadi M, Fagiri MA, Alonazi B (2019) Evolution of Robot-assisted ultrasound-guided breast biopsy systems. J Radiat Res Appl Sci 11(1):89–97
Priester A, Natarajan S, Culjat M (2013) Robotic ultrasound systems in medicine. IEEE Trans Ultrason Ferroelectr Freq Control 60(3):507–523
Boctor EM, Choti MA, Burdette EC, Webster IRJ (2008) Three-dimensional ultrasound-guided robotic needle placement: an experimental evaluation. Int J Med Robot Comput Assist Surg 4(2):180–191
Rousseau F, Hellier P, Barillot C (2005) Confhusius: a robust and fully automatic calibration method for 3D freehand ultrasound. Med Image Anal 9(1):25–38
Hemmerling TM, Taddei R, Wehbe M, Cyr S, Zaouter C, Morse J (2013) First robotic ultrasound-guided nerve blocks in humans using the Magellan system. Anesth Analg 116(2):491–494
Morse J, Terrasini N, Wehbe M, Philippona C, Zaouter C, Cyr S, Hemmerling TM (2014) Comparison of success rates, learning curves, and inter-subject performance variability of robot-assisted and manual ultrasound-guided nerve block needle guidance in simulation. Br J Anaesth 112(6):1092–1097
Neubach Z, Shoham M (2010) Ultrasound-guided robot for flexible needle steering. IEEE Trans Biomed Eng 57(4):799–805
Chevrie J, Shahriari N, Babel M, Krupa A, Misra S (2018) Flexible needle steering in moving biological tissue with motion compensation using ultrasound and force feedback. IEEE Robot Autom Lett 3(3):2338–2345
Xu J, Jia ZZ, Song ZJ, Yang XD, Chen K, Liang P (2010) Three-dimensional ultrasound image-guided robotic system for accurate microwave coagulation of malignant liver tumours. Int J Med Robot Comput Assist Surg 6(3):256–268
Huang Q, Bian GB, Duan X, Zhao HH, Liang P (2010) An ultrasound-directed robotic system for microwave ablation of liver cancer. Robotica 28(2):209–214
Kojcev R, Fuerst B, Zettinig O, Fotouhi J, Lee SC, Frisch B, Taylor R, Sinibaldi E, Navab N (2016) Dual-robot ultrasound-guided needle placement: closing the planning-imaging-action loop. Int J Comput Assist Radiol Surg 11(6):1173–1181
Kobayashi Y, Hong J, Hamano R, Okada K, Fujie M, Hashizume M (2012) Development of a needle insertion manipulator for central venous catheterization. Int J Med Robot Comput Assist Surg 8(1):34–44
Abayazid M, Vrooijink GJ, Patil S, Alterovitz R, Misra S (2014) Experimental evaluation of ultrasound-guided 3D needle steering in biological tissue. Int J Comput Assist Radiol Surg 9(6):931–939
Lim S, Jun C, Chang D, Petrisor D, Han M, Stoianovici D (2019) Robotic transrectal ultrasound-guided prostate biopsy. IEEE Trans Biomed Eng 66(9):2527–2537
Kim C, Chang D, Petrisor D, Chirikjian G, Han M, Stoianovici D (2013) Ultrasound probe and needle-guide calibration for robotic ultrasound scanning and needle targeting. IEEE Trans Biomed Eng 60(6):1728–1734
Abayazid M, Moreira P, Shahriari N, Patil S, Alteroviz R, Misra S (2014) Ultrasound-guided three-dimensional needle steering in biological tissue with curved surfaces. Med Eng Phys 37(1):145–150
Balter M, Chen A, Maguire T, Yarmush M (2017) Adaptive kinematic control of a robotic venipuncture device based on stereo vision, ultrasound, and force guidance. IEEE Trans Ind Electron 64(1):1626–1635
Esteban J, Simson W, Requena Witzig S, Rienmüller A, Virga S, Frisch B, Zettinig O, Sakara D, Ryang Y, Navab N, Hennersperger C (2018) Robotic ultrasound-guided facet joint insertion. Int J Comput Assist Radiol Surg 13(1):895–904
Ashrafian H, Clancy O, Grover V, Darzi A (2017) The evolution of robotic surgery: surgical and anaesthetic aspects. Br J Anaesth 119(suppl_1):i72–i84
Strid JMC, Sauter AR, Ullensvang K, Andersen MN, Daugaard M, Bendtsen MAF, Søballe K, Pedersen EM, Børglum J, Bendtsen T (2017) Ultrasound-guided lumbar plexus block in volunteers; a randomized controlled trial. Br J Anaesth 118(3):430–438
Stoianovici D, Jun C, Lim S, Li P, Petrisor D, Fricke ST, Sharma K, Cleary K (2018) Multi-imager compatible, MR safe, remote center of motion needle-guide robot. IEEE Trans Biomed Eng 65(1):165–177
Shin WH, Kwon DS (2013) Surgical robot system for single-port surgery with novel joint mechanism. IEEE Trans Biomed Eng 60(4):937–944
Choi H, Kwak HS, Lim YA, Kim HJ (2014) Surgical robot for single-incision laparoscopic surgery. IEEE Trans Biomed Eng 61(9):2458–2466
Zettinig O, Frisch B, Virga S, Esposito M, Rienmüller A, Meyer B, Hennersperger C, Ryang Y, Navab N (2017) 3D ultrasound registration-based visual servoing for neurosurgical navigation. Int J Comput Assist Radiol Surg 12:1607–1619
Marhofer P, Harrop-Griffiths W, Kettner SC, Kirchmair L (2010) Fifteen years of ultrasound guidance in regional anaesthesia: part 1. Br J Anaesth 104(5):538–546
O’Donnell BD, O’Sullivan O, Gallagher AG, Shorten GD (2015) Robotic assistance with needle guidance. Br J Anaesth 114(4):708–709
Acknowledgments
We would like to thank Editage (www.editage.cn) for English language editing.
Funding
This study was funded by the National Natural Science Foundation of China (Grant Number 51575343).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Human and animals rights
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards. No animal experiments were performed in this study.
Informed consent
Informed consent was obtained from all participants.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Chen, S., Wang, F., Lin, Y. et al. Ultrasound-guided needle insertion robotic system for percutaneous puncture. Int J CARS 16, 475–484 (2021). https://doi.org/10.1007/s11548-020-02300-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11548-020-02300-1